Complications Associated With the Arterial Puncture Closure Device—Angio-Seal
* To whom correspondence should be addressed. E-mail: a.kadner{at}web.de.
Background Arterial puncture closure devices (APCD) are frequently used after cardiac catheterization. Here, the diagnosis and therapy of femoral artery complications after the use of the Angio-Seal APCD is reported. Patients and methods The Angio-Seal APCD was deployed in 1600 patients undergoing transfemoral catheterization. Results In 7 of 1600 cases (0.4%) vascular complications occurred following Angio-Seal deployment. Diagnosis was made by duplex sonography. Intraoperative findings consisted of a complete occlusion with dissection of the femoral artery in all patients. In 6 cases, the femoral bifurcation had to be reconstructed after endarterectomy. Follow-up is complete with a mean of 6 months. Conclusion The Angio-Seal device should not be used for closure of the superficial femoral artery and in patients with severe arteriosclerosis. The application of arteriography as well as the use of ultrasound-guided puncture is advisable. In all cases, surgical intervention was successful and an adequate therapy for management of complications.